Cryo EM structure of anti-CRISPRs, AcrF1 and AcrF2, bound to type I-F crRNA-guided CRISPR surveillance complex
ELECTRON MICROSCOPY
Refinement
RMS Deviations | |
---|---|
Key | Refinement Restraint Deviation |
f_dihedral_angle_d | 6.731 |
f_angle_d | 0.942 |
f_chiral_restr | 0.05 |
f_bond_d | 0.006 |
f_plane_restr | 0.006 |
Sample |
---|
Anti-CRISPRs AcrF1 and AcrF2 bound P. aeruginosa crRNA-guided CRISPR surveillance(Csy)complex. |
Specimen Preparation | |
---|---|
Sample Aggregation State | PARTICLE |
Vitrification Instrument | HOMEMADE PLUNGER |
Cryogen Name | ETHANE |
Sample Vitrification Details | Sample was applied to poly-L-lysine hydrobromide pre-treated amorphous carbon film coated over a holey grid. Excess sample was blotted with Whatman-1 ... |
3D Reconstruction | |
---|---|
Reconstruction Method | SINGLE PARTICLE |
Number of Particles | 51212 |
Reported Resolution (Å) | 3.4 |
Resolution Method | FSC 0.143 CUT-OFF |
Other Details | Per-frame dose weighting was performed using the RELION particle polishing method. A B-factor of -71 square angstrom was applied to sharpen the final ... |
Refinement Type | |
Symmetry Type | POINT |
Point Symmetry | C1 |
Map-Model Fitting and Refinement | |||||
---|---|---|---|---|---|
Id | 1 | 2 | 3 | ||
Refinement Space | REAL | REAL | REAL | ||
Refinement Protocol | AB INITIO MODEL | RIGID BODY FIT | RIGID BODY FIT | ||
Refinement Target | |||||
Overall B Value | |||||
Fitting Procedure | |||||
Details | Top scoring five models out of two hundred models generated using Rosetta were refined using Phenix and deposited as an ensemble atomic model for the ... | The docked Cas6F structure in the "head" region of the complex was reduced to C-alpha backbone. | Initially two copies were fitted using Chimera and then backbones and side chains were fixed using Coot, followed by real space refinement in Phenix. |
Data Acquisition | |||||||||
---|---|---|---|---|---|---|---|---|---|
Detector Type | GATAN K2 SUMMIT (4k x 4k) | ||||||||
Electron Dose (electrons/Å**2) | 46 |
Imaging Experiment | 1 |
---|---|
Date of Experiment | |
Temperature (Kelvin) | |
Microscope Model | FEI TITAN KRIOS |
Minimum Defocus (nm) | 1200 |
Maximum Defocus (nm) | 2500 |
Minimum Tilt Angle (degrees) | |
Maximum Tilt Angle (degrees) | |
Nominal CS | 2.7 |
Imaging Mode | BRIGHT FIELD |
Specimen Holder Model | FEI TITAN KRIOS AUTOGRID HOLDER |
Nominal Magnification | 29000 |
Calibrated Magnification | |
Source | FIELD EMISSION GUN |
Acceleration Voltage (kV) | 300 |
Imaging Details | Objective astigmatism was corrected at nominal magnification of 29000X, using Thon rings visualized with a K2 camera. |
EM Software | ||
---|---|---|
Task | Software Package | Version |
PARTICLE SELECTION | FindEM | |
IMAGE ACQUISITION | Leginon | |
CTF CORRECTION | CTFFIND3 | 3 |
MODEL FITTING | Coot | 0.8.7 |
INITIAL EULER ASSIGNMENT | RELION | 1.4 |
FINAL EULER ASSIGNMENT | RELION | 1.4 |
CLASSIFICATION | RELION | 1.4 |
RECONSTRUCTION | RELION | 1.4 |
MODEL REFINEMENT | PHENIX | 1.11.1 |
MODEL FITTING | UCSF Chimera | 1.12 |
MODEL REFINEMENT | PHENIX | 1.11.1 |
MODEL FITTING | UCSF Chimera | 1.12 |
MODEL REFINEMENT | PHENIX | 1.11.1 |
Image Processing | ||||
---|---|---|---|---|
CTF Correction Type | CTF Correction Details | Number of Particles Selected | Particle Selection Details | |
NONE | Particles were ctf-corrected during 2D and 3D processing. | 199348 | Template based particle picking was done using FindEM program implemented in Appion processing package. |