Domain organization and conformational plasticity of the G protein effector, PDE6
ELECTRON MICROSCOPY
Sample |
---|
Bovine rod PDE6 holoenzyme in complex with the Fab fragment from the ROS-1 monoclonal antibody |
Sample Components |
Rod cGMP-specific 3',5'-cyclic phosphodiesterase |
monoclonal antibody Fab fragment of ROS-1 |
Specimen Preparation | |
---|---|
Sample Aggregation State | PARTICLE |
Vitrification Instrument | FEI VITROBOT MARK III |
Cryogen Name | ETHANE |
Sample Vitrification Details | Applied 3 uL sample per grid and blotted for 1 second before plunging into liquid ethane (FEI VITROBOT MARK III). |
3D Reconstruction | |
---|---|
Reconstruction Method | SINGLE PARTICLE |
Number of Particles | 12373 |
Reported Resolution (Å) | 11 |
Resolution Method | FSC 0.143 CUT-OFF |
Other Details | A total of 21,100 particles were picked from ice images and CTF corrected using Ctfit. After an initial 3D model was generated as described for PDE6, ... |
Refinement Type | |
Symmetry Type | POINT |
Point Symmetry | C2 |
Map-Model Fitting and Refinement | ||||||
---|---|---|---|---|---|---|
Id | Refinement Space | Refinement Protocol | Refinement Target | Overall B Value | Fitting Procedure | Details |
1 (3DBA) | REAL | RIGID BODY FIT | REFINEMENT PROTOCOL--rigid body DETAILS--Domains w ... | |||
2 (3IBJ) | ||||||
3 (3JWR) | ||||||
4 (3JWR) | ||||||
5 (12E8) | ||||||
6 (12E8) |
Data Acquisition | |||||||||
---|---|---|---|---|---|---|---|---|---|
Detector Type | GATAN ULTRASCAN 4000 (4k x 4k) | ||||||||
Electron Dose (electrons/Å**2) | 15 |
Imaging Experiment | 1 |
---|---|
Date of Experiment | 2011-08-08 |
Temperature (Kelvin) | |
Microscope Model | JEOL 2010F |
Minimum Defocus (nm) | |
Maximum Defocus (nm) | |
Minimum Tilt Angle (degrees) | |
Maximum Tilt Angle (degrees) | |
Nominal CS | 2 |
Imaging Mode | BRIGHT FIELD |
Specimen Holder Model | GATAN LIQUID NITROGEN |
Nominal Magnification | 60000 |
Calibrated Magnification | |
Source | FIELD EMISSION GUN |
Acceleration Voltage (kV) | 200 |
Imaging Details | Parallel beam illumination |
EM Software | ||
---|---|---|
Task | Software Package | Version |
MODEL FITTING | UCSF Chimera | |
RECONSTRUCTION | EMAN |
Image Processing | ||||
---|---|---|---|---|
CTF Correction Type | CTF Correction Details | Number of Particles Selected | Particle Selection Details | |
EMAN ctfit |